35
8.7 Mechanical Specications
• Size: 9.5 x 12.5 x 14.0” (24 x 32 x 35 cm) maximum
• Weight: 21.0 lb (9.5 kg) maximum (not including power cord or shipping box)
• Moisture protection rating: IPX0 (ordinary, per IEC601)
8.8 Environmental Specications
• Operational temperature: 10°C to 35°C
• Storage & Transportation temperature: -40°C to 70°C
• Operation, Storage & Transportation Humidity: 20% to 95% non-condensing
8.9 Fuses
• 100VAC - 240VAC, 50Hz - 60Hz: Replace mains fuses as marked:
4A/250V, T-lag, 5x20mm
• The PMG operates over the AC Voltage range of 100VAC - 240VAC.
•
W
hen replacing fuses, ensure the integrity of the new fuses by inspecting for physical damage that could aect the function of the fuse.
• Use a precision slot drive screwdriver to remove the fuse drawer.
• Replace the fuses, and replace the fuse drawer.
8.10 Line Input Ratings
• 3.5 A @ 100-120 VAC ~ 60 Hz
• 1.7 A @ 220-240 VAC ~ 50 Hz
8.11 Footswitch Specications
• Pneumatic
8.12 Rated Accessory Voltage (for Associated Equipment and Active Accessories)
• 160 V
rms
8.13 Expected Service Life
• 5 Years
8.14 This device is compliant with RoHS Directive 2011/65/EU.