When the reading is stable and close to the selected
buer, the “CFM” tag will blink. Press CFM to conrm
calibration.
After conrming the rst calibration point, the
calibrated value will be displayed on the rst LCD line
and the second expected buer value on the third LCD
line. (i.e. pH 4.01)
Rinse and submerse the pH electrode approximately
3 cm (1¼”) into the second buer solution and stir
gently.
If necessary, press the ARROW keys to select a dierent buer value.
The “
” along with “STIR” tag will be displayed and “WAIT” will blink on the LCD
until the reading is stable. When the reading is stable and close to the selected
buer, the “CFM” tag will blink. Press CFM to conrm calibration.
The calibrated value is then displayed on the rst LCD line and the third expected
buer value on the third LCD line.
After the second calibration point is conrmed, rinse and submerse the pH
electrode approximately 3 cm (1¼”) into the last buer solution and stir gently.
If necessary, press the ARROW keys to select a dierent buer value.
The “ ” along with “STIR” tag will be displayed and “WAIT” will blink on the LCD
until the reading is stable.
When the reading is stable and close to the selected buer, the “CFM” tag will
blink. Press CFM to conrm calibration.
At the end of calibration the instrument displays “SAVING”, stores the calibration
value and returns to normal measurement mode.
The calibration sequence may be reduced to two buer values or a single one.
Press CAL to return to measurement mode after the desired number of buers
have been calibrated.
Note:
• When performing a new calibration or adding to an existing calibration the
rst calibration point will be treated as an oset. See page 24 for details.
• Press CAL after the rst or second calibration point are conrmed and
the instrument will store the calibration data. Then it will return to
measurement mode.
• If the value measured by the instrument is not close to the selected buer,
“WRONG BUFFER” will blink. Check if the correct buer has been used or
selected.