Operational Guide
41
NaCl % GLP
Information
NaCl % Calibration Data in GLP
To view the NaCl% calibration data, press GLP when the instrument is in NaCl%
measurement mode. Use the ARROW keys to scroll through the calibration data.
The instrument will display the calibration temperature and solution.
The cell factor in cm
-1
determined from the calibration
with the current reading.
The salinity coecient determined from the
calibration together with the current reading.
The Solution Standard together with the calibrated
temperature.
The time (hh:mm:ss) of the last calibration together
with the current reading.
The date (yyyy.mm.dd) of the last calibration together
with the current reading.
Calibration Expiration status together with the
current reading:
If disabled, “EXPIRATION WARNING DISABLED” is
displayed.
If enabled, the number of days until the calibration
alarm “CAL DUE” will be displayed. (i.e. “CAL EXPIRES
IN 2 DAYS”)