EasyManua.ls Logo

Hanna Instruments Edge HI 2020 - Dissolved Oxygen Probe Diagram

Hanna Instruments Edge HI 2020
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operational Guide
47
The probe is polarized with a xed voltage of approximately 800 mV between
the cathode and anode. Probe polarization is essential for stable measurements.
With the probe properly polarized, oxygen is continually consumed as it passes
through gas permeable PTFE membrane.
If polarization is interrupted, the electrolyte solution continues to be
enriched with oxygen until it reaches an equilibrium with the surrounding
solution. Whenever measurements are taken with a non-polarized probe, the
measurement will be drifty and inaccurate. The measurement will jump when
the probe is moved.
Note: When not in use and during polarization, use the protective transparent
cap.
Dissolved
Oxygen
Setup
1. Strain Relief
2. Probe Cap
3. PEI Probe Body
4. Temperature Sensor
5. Threads for Membrane Cap
6. Ag/AgCl Anode and Reference
7. Glass Insulator
8. Platinum Cathode
9. O-Ring
10. Disposable Membrane Cap
11. Oxygen Permeable PTFE Membrane
12. Shipping Tube
Dissolved
Oxygen
Probe
Diagram
w ww . . co m
information@itm.com1.800.561.8187

Related product manuals