27
pH
Calibration
Calibration
Messages
• If the buer temperature exceeds the temperature limits of the buer,
“WRONG BUFFER TEMPERATURE” will be displayed.
• Press CLR after entering calibration to clear previous calibrations stored on
probe. “CLEAR CALIBRATION” message will be displayed. Press CFM. The
instrument will return to measurement mode displaying the “CAL DUE”
message.
• If measured value diers more than expected “CHECK ELECTRODE CHECK
BUFFER“ will be displayed. Take appropriate action, cleaning the probe if
necessary or refreshing the pH buer.
pH Buer Temperature Dependence
Calibration buers are aected by temperature. During calibration with standard
pH buers (not Custom), the instrument will display the pH buer value at 25 ºC,
however, it will use the appropriate value for that buer at the temperature of
the buer for the calibration. Immediately after exiting calibration, the buer
will read its value at the temperature of measurement.
TEMP pH BUFFERS
ºC ºF 1.679 3.000 4.010 6.862 7.010 9.177 10.010 12.454
0 32 1.670 3.072 4.007 6.982 7.130 9.459 10.316 13.379
5 41 1.670 3.051 4.002 6.949 7.098 9.391 10.245 13.178
10 50 1.671 3.033 4.000 6.921 7.070 9.328 10.180 12.985
15 59 1.673 3.019 4.001 6.897 7.046 9.273 10.118 12.799
20 68 1.675 3.008 4.004 6.878 7.027 9.222 10.062 12.621
25 77 1.679 3.000 4.010 6.862 7.010 9.177 10.010 12.450
30 86 1.683 2.995 4.017 6.851 6.998 9.137 9.962 12.286
35 95 1.688 2.991 4.026 6.842 6.989 9.108 9.919 12.128
40 104 1.693 2.990 4.037 6.837 6.983 9.069 9.881 11.978
45 113 1.700 2.990 4.049 6.834 6.979 9.040 9.847 11.834
50 122 1.707 2.991 4.062 6.834 6.978 9.014 9.817 11.697
55 131 1.715 2.993 4.076 6.836 6.979 8.990 9.793 11.566
60 140
1.724 2.995 4.091 6.839 6.982 8.969 9.773 11.442
65 149 1.734 2.998 4.107 6.844 6.987 8.948 9.757 11.323
70 158 1.744 3.000 4.123 6.850 6.993 8.929 9.746 11.211
75 167 1.755 3.002 4.139 6.857 7.001 8.910 9.740 11.104
80 176 1.767 3.003 4.156 6.865 7.010 8.891 9.738 11.003
85 185 1.780 3.002 4.172 6.873 7.019 8.871 9.740 10.908
90 194 1.793 3.000 4.187 6.880 7.029 8.851 9.748 10.819
95 203 1.807 2.996 4.202 6.888 7.040 8.829 9.759 10.734
The Calibration Check™ feature may ag diagnostic messages during a
calibration. As electrode aging is normally a slow process, substantial changes
from previous calibrations are likely due to a temporary problem with the probe
or buers that can be addressed easily. These messages are seen in Standard
and Basic modes.
Wrong Buer
This message appears when the dierence between
the pH reading and the value of the selected buer
is too great. If this error message is displayed, check
if you have selected the proper calibration buer and
have poured the desired buer.