EasyManuals Logo
Home>Hanna Instruments>Measuring Instruments>Edge HI 2020

Hanna Instruments Edge HI 2020 User Manual

Hanna Instruments Edge HI 2020
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
Operational Guide
44
Note:
These are for determining salinity as they relate to general oceanographic
use.
Practical salinity and the Natural Seawater require a conductivity
calibration.
NaCl % requires a calibration in HI 70371 standard.
PSU - Practical Salinity Units
The practical salinity (S) of seawater relates the ratio of electrical conductivity
of a normal seawater sample at 15 ºC and 1 atmosphere to a potassium chloride
solution (KCl) with a mass of 32.4356 g/Kg water at the same temperature and
pressure.
Under these conditions the ratio is equal to 1, and S=35. The practical salinity
scale may be applied to values to through 42 PSU at temperatures between -2
to 35 ºC.
According to the denition, salinity of a sample in PSU (practical salinity units) is
calculated using the following formula:
NaCl % Percent Scale
The NaCl % scale is an older salinity scale used for seawater salinity. In this scale
100% salinity is equivalent to roughly 10% solids. High percentages were made
by evaporation. To display NaCl in % units, enter SETUP and select NaCl% unit.
Press the RANGE key until “NaCl%” is displayed on the LCD. The instrument will
display the NaCl% reading on the rst LCD line and the temperature reading on
the second LCD line.
Salinity
Measurements
f(t)=
T-15
1 + 0.0162•(t-15)
a
0
= 0.008
a
1
= -0.1692
a
2
= 25.3851
a
3
= 14.0941
a
4
= -7.0261
a
5
= 2.7081
b
0
= 0.0005
b
1
= -0.0056
b
2
= -0.0066
b
3
= -0.0375
b
4
= 0.0636
b
5
= -0.01442
c
0
= 0.008
c
1
= 0.0005
X = 400R
T
Y = 100R
T
R
T
- ratio of sample conductivity to standard conductivity at Temp =(T)
C
T
(sample) - uncompensated conductivity at T ºC;
C(35,15) =42.914µS/cm - the corresponding conductivity of KCI solution
containing a mass of 32.4356 g KCl/1 Kg solution
r
T
- Temperature compensation polynomial
w ww . . co m
information@itm.com1.800.561.8187

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments Edge HI 2020 and is the answer not in the manual?

Hanna Instruments Edge HI 2020 Specifications

General IconGeneral
BrandHanna Instruments
ModelEdge HI 2020
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals