Operational Guide
26
When the reading is stable and close to the selected
bu er, the “CFM” tag will blink. Press CFM to con rm
calibration.
After con rming the rst calibration point, the
calibrated value will be displayed on the rst LCD line
and the second expected bu er value on the third LCD
line. (i.e. pH 4.01)
Rinse and submerse the pH electrode approximately
3 cm (1¼”) into the second bu er solution and stir
gently.
If necessary, press the ARROW keys to select a di erent bu er value.
The
along with “STIR” tag will be displayed and “WAIT” will blink on the LCD
until the reading is stable. When the reading is stable and close to the selected
bu er, the “CFM” tag will blink. Press CFM to con rm calibration.
The calibrated value is then displayed on the rst LCD line and the third expected
bu er value on the third LCD line.
After the second calibration point is con rmed, rinse and submerse the pH
electrode approximately 3 cm (1¼”) into the last bu er solution and stir gently.
If necessary, press the ARROW keys to select a di erent bu er value.
The
along with “STIR” tag will be displayed and “WAIT” will blink on the LCD
until the reading is stable.
When the reading is stable and close to the selected bu er, the “CFM” tag will
blink. Press CFM to con rm calibration.
At the end of calibration the instrument displays “SAVING”, stores the calibration
value and returns to normal measurement mode.
The calibration sequence may be reduced to two bu er values or a single one.
Press CAL to return to measurement mode after the desired number of bu ers
have been calibrated.
Note:
When performing a new calibration or adding to an existing calibration the
rst calibration point will be treated as an o set. See page 24 for details.
Press CAL after the rst or second calibration point are con rmed and
the instrument will store the calibration data. Then it will return to
measurement mode.
If the value measured by the instrument is not close to the selected bu er,
“WRONG BUFFER” will blink. Check if the correct bu er has been used,
or clean the electrode by following the Cleaning Procedure. If necessary,
change the bu er or the electrode.
pH
Calibration