11
Reference
Filling Hole
Reference
Filling Hole
Sensitive
Wire
Reference
Wire
Reference
Junction
Glass
Bulb
Glass
Bulb
Reference
Junction
Reference
Junction
Reference
Junction
Platinum or
Gold tip
Platinum or
Gold tip
Reference
Wire
Reference
Wire
Sensitive
Wire
Glass Body
pH Electrode
Plastic Body
pH Electrode
Plastic Body
ORP Electrode
Glass Body
ORP Electrode
Reference
Wire
ELECTRODE CONDITIONING
AND MAINTENANCE
PREPARATION PROCEDURE
Remove the protective cap.
DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal
with electrodes and they will disappear when rinsed with water.
During transport tiny bubbles of air may form inside the glass bulb
affecting proper functioning of the electrode. These bubbles can be
removed by "shaking down" the electrode as you would do with a
glass thermometer.
If the bulb and/or junction is dry, soak the electrode in HI 70300
storage solution for at least one hour.
Note: To prevent damage to the electrode, remove the pH
electrode from the solution before turning the meter off.
If the meter is OFF, detach the electrode from the meter
before immersing the electrode in the storage solution.