EasyManuals Logo

Hanna Instruments HI83303 User Manual

Hanna Instruments HI83303
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
35
AMMONIA LOW RANGE
8.3. AMMONIA LOW RANGE
SPECIFICATIONS
Range 0.00 to 3.00 mg/L (as NH
3
‑N)
Resolution 0.01 mg/L
Accuracy ±0.04 mg/L ±4% of reading at 25 °C
Light Source LED with narrow band interference filter @ 420 nm
Method Adaptation of the ASTM Manual of Water and Environmental Technology,
D1426 Nessler method.
REQUIRED REAGENTS
Code Description Quantity
HI93700A‑0 Ammonia Low Range Reagent A 4 drops
HI93700B‑0 Ammonia Low Range Reagent B 4 drops
REAGENT SETS
HI93700‑01 Reagents for 100 tests
HI93700‑03 Reagents for 300 tests
For other accessories see page 87.
MEASUREMENT PROCEDURE
• Select the Ammonia LR method using the procedure described in the
Method Selection section (see page 17).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
10 mL