EasyManuals Logo

Hanna Instruments HI83303 User Manual

Hanna Instruments HI83303
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #69 background imageLoading...
Page #69 background image
69
OXYGEN, DISSOLVED
8.16. OXYGEN, DISSOLVED
SPECIFICATIONS
Range 0.0 to 10.0 mg/L (as O
2
)
Resolution 0.1 mg/L
Accuracy ±0.4 mg/L ±3% of reading at 25 °C
Light Source LED with narrow band interference filter @ 420 nm
Method Adaptation of the Standard Methods for the Examination of Water and
Wastewater, 18th edition, Azide modified Winkler method.
REQUIRED REAGENTS
Code Description Quantity
HI93732A‑0 Dissolved Oxygen Reagent A 5 drops
HI93732B‑0 Dissolved Oxygen Reagent B 5 drops
HI93732C‑0 Dissolved Oxygen Reagent C 10 drops
REAGENT SET
HI93732‑01 Reagents for 100 tests
HI93732‑03 Reagents for 300 tests
For other accessories see page 87.
MEASUREMENT PROCEDURE
• Select the Oxygen (dissolved) method using the procedure
described in the Method Selection section (see page 17).
• Fill one 60 mL glass bottle completely with the unreacted
sample.
• Replace the cap and ensure that a small part of the sample
spills over.
• Remove the cap and add 5 drops of HI93732A‑0 and 5 drops of
HI93732B‑0.
• Add more sample, to fill the bottle completely. Replace the cap
and ensure that a part of the sample spills over.
Note: This ensures no air bubbles have been trapped inside
the bottle. Trapped air bubbles could alter readings.
• Invert the bottle several times until the sample turns orange‑
yellow and a flocculating agent appears.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI83303 and is the answer not in the manual?

Hanna Instruments HI83303 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI83303
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals