EasyManua.ls Logo

Hanna Instruments HI83303 - Page 70

Hanna Instruments HI83303
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
70
OXYGEN, DISSOLVED
• Let the sample stand for approximately 2 minutes to allow flocculating
agent to settle.
• When the upper half of the bottle is clear, add 10 drops of
HI93732C‑0 Dissolved Oxygen Reagent C.
• Replace the cap and invert the bottle until the settled flocculating
agent dissolves completely. The sample is ready for measurement
when it is yellow and completely clear.
• Fill the first cuvette (#1) with 10 mL of the unreacted sample
(up to the mark), and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Fill second cuvette (#2) with 10 mL of the reacted sample
(up to the mark) and replace the cap.
10 mL
#1

Related product manuals