EasyManuals Logo

Hanna Instruments HI83303 User Manual

Hanna Instruments HI83303
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
33
ALKALINITY, MARINE
10 mL
8.2. ALKALINITY, MARINE
SPECIFICATIONS
Range 0 to 300 mg/L (as CaCO
3
)
Resolution 1 mg/L
Accuracy ±5 mg/L ±5% of reading at 25 °C
Light Source LED with narrow band interference filter @ 610 nm
Method Colorimetric Method.
REQUIRED REAGENTS
Code Description Quantity
HI755S Alkalinity Reagent 1 mL
REAGENT SETS
HI755‑26 Reagents for 25 tests
For other accessories see page 87.
MEASUREMENT PROCEDURE
• Select the Alkalinity Marine method using the procedure described in
the Method Selection section (see page 17).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI83303 and is the answer not in the manual?

Hanna Instruments HI83303 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI83303
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals