COPPER HIGH RANGE
8.11. COPPER HIGH RANGE
SPECIFICATIONS
Range 0.00 to 5.00 mg/L (as Cu)
Resolution 0.01 mg/L
Accuracy ±0.02 mg/L ±4% of reading at 25 °C
Light Source LED with narrow band interference filter @ 575 nm
Method Adaptation of the EPA method.
REQUIRED REAGENTS
Code Description Quantity
HI93702‑0 Copper High Range Reagent 1 packet
REAGENT SETS
HI93702‑01 Reagents for 100 tests
HI93702‑03 Reagents for 300 tests
For other accessories see page 87.
MEASUREMENT PROCEDURE
• Select the Copper HR method using the procedure described in the
Method Selection section (see page 17).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
10 mL