EasyManua.ls Logo

Hanna Instruments HI83303 - Copper High Range

Hanna Instruments HI83303
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
57
COPPER HIGH RANGE
8.11. COPPER HIGH RANGE
SPECIFICATIONS
Range 0.00 to 5.00 mg/L (as Cu)
Resolution 0.01 mg/L
Accuracy ±0.02 mg/L ±4% of reading at 25 °C
Light Source LED with narrow band interference filter @ 575 nm
Method Adaptation of the EPA method.
REQUIRED REAGENTS
Code Description Quantity
HI93702‑0 Copper High Range Reagent 1 packet
REAGENT SETS
HI93702‑01 Reagents for 100 tests
HI93702‑03 Reagents for 300 tests
For other accessories see page 87.
MEASUREMENT PROCEDURE
• Select the Copper HR method using the procedure described in the
Method Selection section (see page 17).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
10 mL

Related product manuals