8.14. NITRITE LOW RANGE
SPECIFICATIONS
Range 0 to 600 µg/L (as NO
2
‑N)
Resolution 1 µg/L
Accuracy ±20 µg/L ±4% of reading at 25 °C
Light Source LED with narrow band interference filter @ 466 nm
Method Adaptation of the EPA Diazotization method 354.1.
REQUIRED REAGENTS
Code Description Quantity
HI93707‑0 Nitrite Low Range Reagent 1 packet
REAGENT SETS
HI93707‑01 Reagents for 100 tests
HI93707‑03 Reagents for 300 tests
For other accessories see page 87.
MEASUREMENT PROCEDURE
• Select the Nitrite LR method using the procedure described
in the Method Selection section (see page 17).
• Fill the cuvette up to the mark with 10 mL of unreacted
sample (up to the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
10 mL
NITRITE LOW RANGE