EasyManuals Logo

Hanna Instruments HI83303 User Manual

Hanna Instruments HI83303
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #52 background imageLoading...
Page #52 background image
52
8.9. CHLORINE, TOTAL
SPECIFICATIONS
Range 0.00 to 5.00 mg/L (as Cl
2
)
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading at 25 °C
Light Source LED with narrow band interference filter @ 525 nm
Method Adaptation of the EPA DPD method 330.5.
REQUIRED REAGENTS
POWDER:
Code Description Quantity
HI93711‑0 Total Chlorine Reagent 1 packet
LIQUID:
Code Description Quantity
HI93701A‑T Total Chlorine Reagent A 3 drops
HI93701B‑T Total Chlorine Reagent B 3 drops
HI93701C‑T Total Chlorine Reagent C 1 drop
REAGENT SETS
HI93701‑T Reagents for 300 tests (liquid)
HI93711‑01 Reagents for 100 total tests (powder)
HI93711‑03 Reagents for 300 total tests (powder)
For other accessories see page 87.
MEASUREMENT PROCEDURE
• Select the Chlorine (Total) method using the procedure described in
the Method Selection section (see page 17).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
10 mL
CHLORINE, TOTAL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI83303 and is the answer not in the manual?

Hanna Instruments HI83303 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI83303
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals