53
CHLORINE, TOTAL
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
POWDER REAGENT PROCEDURE
• Add 1 packet of HI93711‑0 Total Chlorine Reagent. Replace the
cap and shake gently for 20 seconds.
• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively,
wait for 2 minutes and 30 seconds and press Read. When the timer ends the meter will perform
the reading. The instrument displays the results in mg/L of chlorine (Cl
2
).
LIQUID REAGENT PROCEDURE
• To an empty cuvette add 3 drops of HI93701A‑T
Total Chlorine Reagent A, 3 drops of HI93701B‑T
Total Chlorine Reagent B, and 1 drop of
HI93701C‑T Total Chlorine Reagent C. Swirl gently
to mix.