Operating Data Logger
USER’S MANUAL 24
The user can select Auto measure and Manual measure. Voltage range can be set 60.00mV,
600.0mV, 6.000V, 60.00V, 600.0V.
2.2 Making Current Measurement
1. Making a DC Current Measurement
To measure a DC current which is less than 600mA, follow these steps:
1) Click the A key and then DC appears on the screen. The unit on the main reading screen is
mA. Click A key to switch the measurement between mA and A. The default is 600mA. Click
Auto key to switch the measurement range from 60mA to 600mA.
2) Connect the black lead into the COM banana jack input and the red lead into the mA banana
jack input.
3) Connect the red and black leads to the measured points and the DC current value of measured
points will be displayed on the screen.
Connect Data Logger probes(HT325) with Hantek365 as illustrated in the following figure:
Figure 2-5
Then, the following figure will be displayed on the screen:
Figure 2-6 DC current Measurement for 600mA