Safety requirements 
    EN 
Tablet1000 Quick guide                                                                  Copyright © Qingdao Hantek Electronics Co., LTD 
4 
Installation (Overvoltage) Class This product is powered by a main power supply 
conforming to Installation (overvoltage) Class II. 
Warning: 
Ensure that no overvoltage (such as voltage caused by lightning) reaches the 
product. Otherwise, the operator may suffer electric shocks. 
Installation (overvoltage) class definition 
Class I refers to the signal level, which is applicable to the measuring terminal of the 
instrument connected to the source circuit. Measures have been taken to limit the 
instantaneous voltage to the corresponding low level. 
Class II refers to the local distribution level, which is applicable to instruments connected 
to the mains (AC) supply. 
Pollution level 
Level 2 
Pollution level definition 
⚫  Level 1: No pollution, or only dry non-conductive pollution. This pollution level has 
no impact. For example: clean room or air-conditioned office environment. 
⚫  Level 2: Generally, only dry non-conductive pollution. Temporary conduction due to 
condensation may sometimes occur. For example: general indoor environment. 
⚫  Level 3: Conductive pollution, or dry non-conductive pollution becoming conductive 
due to condensation. For example, an outdoor environment with a canopy. 
⚫  Level 4: Permanent conductive pollution through conductive dust, rain, or snow. 
For example: outdoor places. 
Security level 
Level 1 - Grounded products 
1.5  Maintenance and cleaning   
Maintenance: 
Do not make the LCD exposed to direct sunlight for a long time when placing the 
oscilloscope. 
Cleaning: 
Regularly check the oscilloscope and probe according to the requirements of the 
operating conditions, please follow the following steps to clean the outer surface of the 
instrument: 
1) Using a lint-free cloth to remove the dust from the outside of the oscilloscope and