EasyManua.ls Logo

HBM AE101 - Page 20

HBM AE101
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
21
Clip
HBMA0114-6.3 en
Example 3
Force transducer 100 N ¢ 2 m V/V; measuring range wanted: 60 N ¢ 10 V
60 N
100 N
10 V
10 V
2mVńV + 1.2 mVńV
Load transducer with 60 N.
Use S21 to select appropriate interval as given in Tab. 4.2 (page 19) (switch
on 4 + 7), and use P22 to adjust the output voltage to 10 V exactly. The mea-
suring range can also be adjusted with no transducer connected by means of
a calibration unit (e.g. K 3607) that simulates the transducer signal.
Additive calibration signal
The measuring range can also be adjusted using the additive calibration sig-
nal (0.2 mV/V).
Example 4
Force transducer 100 N ¢ 2mV/V; measuring range wanted: 60 N ¢ 10 V
60 N
100 N
10 V
10 V
2mVńV + 1.2 mVńV
Calculate nominal measurement value, execute zero balance (see above) and
use S21 to select the corresponding measuring-range interval. Connect addi-
tive calibration signal (set S25 to CAL) and use P22 to adjust with the trans-
ducer under no load. The amplifier has now been calibrated. Adjust S25 to
MEAS.
0.2 mVńV
1.2 mVńV
10 V + 1.667 V
Synchronization
For synchronization of several CF amplifiers, one master device must be de-
fined by adjusting S24 to MASTER. The other devices must be set to SLAVE
using S24. Then, interconnect terminals 8. Synchronization is also possible
with AE501 measuring amplifiers.

Related product manuals