AED9301B cable connection via PG glands
23
AED9301B HBM
3.6 AED9301B cable connection via PG glands
Only a connecting cable with a screen grounded on two sides should be used as the inter-
connecting cable between the AED 9301B and the transducer or Profibus DP. On the AED
side, bring the screen extensively into contact at the PG gland. On the other side, connect
the screen extensively to ground (housing). If there are vast differences between the ground
potential of the AED9301B and its partner device, a potential equalization line must be pro-
vided in addition.
Fig. 3.6-1: Cable connection via a PG gland