Helmer Scientic i.Series
®
and Horizon Series
™
Freezer - Upright Operation Manual
360369/A 12
i.Series Operation
3.5 Active Alarms
Home screen with active alarm
Table 1. i.Series Active Alarms
Alarm Description
Primary Monitor Probe High Temp Primary monitor probe reading is above high temperature alarm setpoint
Primary Monitor Probe Low Temp Primary monitor probe temperature reading is below low temperature alarm setpoint
Primary Monitor Probe Failure Primary monitor probe not functioning properly
Secondary Monitor Probe High Temp
(if installed)
Secondary monitor probe reading is above high temperature alarm setpoint
Secondary Monitor Probe Low Temp
(if installed)
Secondary monitor probe temperature reading is below low temperature alarm setpoint
Secondary Monitor Probe Failure
(if installed)
Secondary monitor probe not functioning properly
Control Probe Failure Control probe not functioning properly
Evap Defrost Probe Failure Evaporator defrost probe not functioning properly
Compressor High Temperature Compressor temperature reading is above high temperature alarm setpoint
Compressor Probe Failure Compressor probe not functioning properly
Power Failure Power to unit has been disrupted
Door Open Door is open beyond user-specied duration
Low Battery Rechargeable battery voltage is low
No Battery Battery is not connected
Communication Failure Messages
1, 2, 3
1 Communication lost between i.C³ display board and control board
2 Communication lost between i.C³ display board and internal system memory
3 Corrupt database