8 ProductSpecications
8.1 Operating Standards
These units are designed to operate under the following environmental conditions:
Indoor use only
Altitude (maximum): 2000 m
Ambient temperature range: 15 °C to 32 °C
Relative humidity (maximum for ambient temperature): 80% for temperatures up to 31 °C, decreasing linearly to 50% at 40 °C
Temperature control range: -15 °C to -30 °C
Table5.ElectricalSpecications(LaboratoryandPlasmaStorage)
Model 120 / 125
Input Voltage and
Frequency
115 V, 60 Hz;
230 V, 50 Hz;
208/230 V, 60 Hz
Voltage Tolerance ±10%
Circuit Breakers
(230 V only)
12.0 A (quantity 2)
Current Draw 8.5 A (115 V, 60 Hz)
3.8 A (230 V, 50 Hz)
4.3 A (208/230 V, 60 Hz)
Power Source Grounded outlet, meeting national electric code (NEC) in
the U.S. and local electrical requirements in all locations.
Remote Alarm
Capacity
i.Series: 0.5 A at 30 V (RMS); 1.0 A at 24 V (DC)
Horizon Series: 0.25 A at 30 V (RMS); 0.25 A at 60 V (DC)
CAUTION
The interface on the remote alarm monitoring system is intended for connection to the end user’s central alarm
system(s) that uses normally-open or normally-closed dry contacts.
If an external power supply exceeding 30 V (RMS) or 60 V (DC) is connected to the remote alarm monitoring system’s
circuit, the remote alarm will not function properly; may be damaged; or may result in injury to the user.
Notes
Add 0.375” (10 mm) to width for optional access control.
Maximum load per drawer, shelf or basket - 100 lbs (46 kg).
Table6.LaboratoryFreezerSpecications
Model Voltage Code Amps
Cu. Ft/
Liters Cabinet Door Shelves
Dimensions W x H x D in. (mm)
Exterior
Net Wt.
lbs (kg)
iLF120
115 V 60 Hz 8.5
20
(566)
Upright
Single hinged
solid
4
30.75 x 80 x 32.5
(782 x 2032 x 826)
443
(201)
230 V 50 Hz 3.8
208/230 V 60 Hz 4.3
HLF120
115 V 60 Hz 8.5
20
(566)
Upright
Single hinged
solid
4
30.75 x 80 x 32.5
(782 x 2032 x 826)
440
(200)
230 V 50 Hz 3.8
208/230 V 60 Hz 4.3
iLF125
115 V 60 Hz 8.5
25
(708)
Upright
Single hinged
solid
4
30.75 x 80 x 38.5
(782 x 2032 x 978)
481
(219)
230 V 50 Hz 3.8
208/230 V 60 Hz 4.3
HLF125
115 V 60 Hz 8.5
25
(708)
Upright
Single hinged
solid
4
30.75 x 80 x 38.5
(782 x 2032 x 978)
478
(217)
230 V 50 Hz 3.8
208/230 V 60 Hz 4.3