6 Horizon Series™ Operation
6.1 Initial Power Up
1. Plug the power cord into a grounded outlet that meets the electrical requirements on the product specication label.
2. Turn the AC power switch ON.
3. Install 9 V back-up battery located on top of unit.
4. Press Down Arrow (Mute) if high temperature alarm sounds.
Notes
For models equipped with optional Access Control, the back up battery is turned ON with a key switch.
If an alarm condition other than High Temperature occurs, refer to the service manual for troubleshooting.
Monitor and Control Panel
6.2 Display Minimum and Maximum Monitor Temperature Recordings
Notes
This feature is standard on Horizon Series™ models with serial numbers of 2015494 or higher. Some exceptions may
exist. For conrmation on your unit, please contact Helmer Technical Service.
This feature only applies to the Monitor temperature probe.
Units which do not include the minimum and maximum recording feature will not display ˚C or ˚F when entering the
program mode.
The minimum and maximum recording feature allows the user to view a minimum temperature occurrence and a maximum
temperature occurrence within a given period of time. The timer provides a time reference in which those temperatures occurred.
View Minimum Temperature Recording
1. Press and hold the Down Arrow button for 1 second and listen for a single beep.
2. The display will alternate between LO and a valid temperature value ve (5) times followed by a single beep to indicate
exit back to the temperature display.
View Maximum Temperature Recording
1. Press and hold the Up Arrow button for 1 second and listen for a single beep.
2. The display will alternate between HI and a valid temperature value ve (5) times followed by a single beep to indicate
exit back to the temperature display.
LED Display
High & Low Temperature Lamps
Up & Down Arrows
SET Button
Control Lamp
Monitor Lamp
SEL Button
Horizon Series Operation