EasyManua.ls Logo

Heraeus UT6760 - 6. Appendix Table

Default Icon
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6. Appendix - Table 3, from DIN-Standard 12 880
Safety devices against malfunction of the temperature control circuit.
Classification of heating ovens according to DIN-Standard 12 880 Part 1, table 3.
The class describes the type and scope of protection, as well as the safety measures
required to deal with a simple defect in the temperature control circuit.
Class
0
1
2
3
3.1
3.2
3.3
Purpose
-
Protection
of oven
Protection
of oven,
environment
and mate-
rial being
treated.
Scope
-
In
the
event
of a
defect,
no
danger
emanates from
the
oven.
In
the
event
of a
defect,
no
danger
emanates from
the
oven
or
from
the ma-
terial being treated.
In
the
event
of a
defect,
the
material
being treated
is
pro-
tected against over-
heating and/or under-
cooling (e.g. incuba-
tor)
Safety device
ace.
to
Sub-
section
5.5
-
Temperature
11 miter
or
thermal fuse
Adjustable
temperature
limit cut-out
Adjustable
temperature
limit
con-
troller,
coming into
action
in
case
of:
excess
temperature
Insufficient
temperature
excess
temperature
and
insufficent
temperature
Safety measures
Only attended opera-
tion*) with harmless
materials
1s
permit-
ted;
or
overheating
is excluded
by
con-
structional measures.
Special safety meas-
ures depending
on
the intended
use.
*) If operation is supervised, the oven must be checked at reasonably short intervals to
make sure it works correctly.