EasyManua.ls Logo

Hettich EBA 200 S - Querying Operating Hours and Centrifugation Runs

Hettich EBA 200 S
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AB1800en_SARev.: 03 / 11.2023
28 / 39
Software operation
Resetting the cycle counter
The centrifuge is equipped with a cycle counter. The cycle counter counts
the run cycles (centrifugation cycles). The remaining number of run cycles
(centrifugation runs) is displayed briey after each centrifugation run.
If the maximum permissible number of rotor run cycles entered is
exceeded,
‘Cycles passed’
is displayed after each start of a centrifugation
run. The centrifugation run must be restarted. The rotor must be replaced
with a new one.
Once the rotor has been replaced, the cycle counter must be reset to
‘0’
.
The cycle counter must be reset to
‘0’
after installing a new rotor.
1. Press and hold the
[SELECT]
button.
🡆
‘*MACHINE MENU*’
is displayed after 8 seconds.
2. Press the
[SELECT]
button repeatedly until
‘-> Time & Cycles’
is
displayed.
3. Press the
[START/PULSE]
button.
4. Press the
[SELECT]
button repeatedly until
‘Cyc sum=...’
is displayed.
5. Press the
[RCF]
button.
6.
Press the
[t
]
button.
🡆 The number of run cycles completed is reset to
‘0’
.
7. Press the
[START/PULSE]
button.
🡆
‘Store cycles...’
is displayed.
8. Press the
[STOP/OPEN]
button twice to exit the
‘-> Time & Cycles’
menu
or
Press the
[STOP/OPEN]
button three times to exit the
‘*MACHINE
MENU*’
.
7.2.3 Querying operating hours and centrifugation runs
The operating hours are divided into internal and external operating hours.
Internal operating hours: Total time for which the device has been
switched on.
External operating hours: Total time of centrifugation runs to date.
The rotor is stationary.
1. Press and hold the
[SELECT]
button.
🡆
‘*MACHINE MENU*’
is displayed after 8 seconds.
2. Press the
[SELECT]
button repeatedly until
‘-> Time & Cycles’
is
displayed.
3. Press the
[START/PULSE]
button.
🡆
‘TimeExt=’
is displayed.
TimeExt: External operating hours
4. Press the
[SELECT]
button.
🡆
‘TimeInt=’
is displayed.
TimeInt: Internal operating hours
5. Press the
[SELECT]
button.
🡆
‘Starts=’
is displayed.
Starts: Number of all centrifugation runs