EasyManua.ls Logo

Hettich MIKRO 200 - Short-term centrifugation; Quick stop function; Software operation; Centrifugation parameters

Hettich MIKRO 200
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AB2400en_SARev.: 03 / 11.2023
26 / 40
Software operation
3. Ramp-down takes place with the selected brake level after the time
has elapsed or if the centrifugation run is cancelled.
🡆 The brake level is displayed.
An audible signal sounds when the rotor comes to a standstill.
‘OPEN’
‘OEFFNEN’
is displayed.
The right side of the
[STOP/OPEN]
button lights up when the
centrifuge is in ramp-down.
The left side of the
[STOP/OPEN]
button lights up when the rotor is
at a standstill.
The light on the
[START/IMPULS]
button and the right side of the
[STOP/OPEN]
button go out.
6.5.3
Short-term centrifugation
Personnel:
â– 
Trained user
1. Press and hold the
[START/IMPULS]
button.
🡆 The
[START/IMPULS]
button lights up during the centrifugation
run.
Timing starts at 00:00.
The rotor speed or the RCF value, the temperature in the centri-
fuging chamber (only for centrifuges with cooling) and the elapsed
time are displayed during the centrifugation run.
2. Release the
[START/IMPULSE]
button to end the centrifugation run.
🡆 Ramp-down takes place with the set brake level. The brake level is
displayed.
An audible signal sounds when the rotor comes to a standstill.
‘OPEN’
‘OEFFNEN’
is displayed.
6.6
Quick stop function
Personnel:
â–  Trained user
Press the
[STOP/OPEN]
button twice.
🡆
Ramp-down with brake level "9" (shortest ramp-down time) is dis-
played and executed.
7
Software operation
7.1 Centrifugation parameters
7.1.1 Relative centrifugal force, RCF
The relative centrifugal force RCF is dependent on the speed and the centri-
fuging radius.
The relative centrifugal force RCF is stated as a multiple of the acceleration
due to gravity (g).
The relative centrifugal force RCF is a dimensionless numerical value and is
used to compare the separation and sedimentation performance.

Table of Contents