Software operation
5. Turn to the right with the
[Rotary knob]
.
🡆
‘ROTORCHG1’
is displayed.
Internal operating hour of the last rotor change
6. Turn to the right with the
[Rotary knob]
.
🡆
‘ROTORCHG2’
is displayed.
Internal operating hour of the penultimate rotor change
7. Turn to the right with the
[Rotary knob]
.
🡆
‘OPhoursCHG’
is displayed.
Internal operating hour of the last operating hours change
8. Turn to the right with the
[Rotary knob]
.
🡆
‘IMBALCHG’
is displayed.
Internal operating hour of the last imbalance cut-off change
9. Turn to the right with the
[Rotary knob]
.
🡆
‘OffsetCHG’
is displayed.
Internal operating hour of the last offset adjustment
10. Press the STOP/OPEN button to exit the menu.
7.4.2
Querying operating hours
The rotor is stationary.
1. Press and hold the
[SELECT]
button.
🡆
‘SOUND/BELL’
is displayed after 8 seconds.
2. Press the
[SELECT]
button repeatedly until
‘CONTROL:’
is displayed.
🡆
‘CONTROL:’
and the operating hours are displayed.
3. Press the
[STOP/OPEN]
button to exit the menu.
7.4.3
Audible signal
7.4.3.1
General
The audible signal sounds:
■ after a problem occurs in the 2s interval.
■ after completion of the centrifugation run and rotor standstill in the 30 s
interval.
Opening the lid or pressing any button stops the audible signal.
7.4.3.2
Setting an audible signal
1.
Press and hold the
[SELECT]
button.
🡆
‘SOUND / BELL ON’
or
‘SOUND / BELL OFF’
is displayed after 8
seconds.
2.
Use
[Rotary knob]
to set
‘OFF’
or
‘ON’
.
OFF = audible signal disabled
ON = audible signal enabled
3.
Press the
[START/IMPULS]
button.
🡆
The setting is stored.
‘*** ok ***’
is displayed briey.