When carrying out centrifuging runs
in
series and
at
the same speed, only
centrifuging times must be reset.
When the centrifuge has been loaded unevenly, its drive will cut out
immediately. The rotation indicator will flash every 0.5 seconds.
-
When rotor has come to a stop, open cover and load rotor evenly.
Impulse Operation
Set timer to "0".
Press start key and hold for as long as this centrifuging run is to continue. Rotor will
turn at the speed preselected via its speed selector.
Stopping the Centrifuge
At the end
of a
preset centrifuging time, the timer will shut down the centrifuge. The
rotor will be braked and slowed down until its stops.
The centrifugation process can be interrupted at any time by resetting timer or
speed selector to "0".
When rotor has come to a stop, open cover and take out centrifugation material.
Set mains switch to position "0".
Changing
a
Rotor
Removal
Open cover
« Undo rotor tightening nut, using spanner wrench, and turn until lift-off pressure point
is reached. When this lift-off pressure point has been overcome, the rotor will pome
off its motprshaft cone. Turn tightening nut until rotor can be lifte^pff !J§ motpTShaft.
Lift rotor off its motor shaft.