1. After adjusting the width of the bed, a beep continues to sound. Why?
The Compella™ Bed can be adjusted to either 40" or 50". The continuous beep is intended to
alert you the bed is not fully adjusted. Continue to press and hold the expand or “retract
width adjustment until you hear a single beep confirmation. This will let you know the bed is
fully adjusted.
2. Why will the bed width not adjust when I press the “Width Expand” or Width Retract”
button?
Ensure both power cords for the bed and surface are plugged into the power. Powered width
adjustment controls only work when connected to power.
If, after verifying bed and surface are connected to power, and the bed will still not adjust,
verify the communication cable is securely connected between the surface blower box and
the frame.
3. When trying to weigh my patient, a beep sounds and the raise bed” indicator flashes,
and I’m unable to weigh the patient. Why?
If the bed is equipped with IntelliDrive® XL Transport System, you cannot weigh a patient
when the bed is in transport mode. To remove the bed from the transport mode, on the
caregiver control pod, press and hold the “raise bed” control until you hear a single beep
confirmation. You can now weigh the patient.
4. When trying to arm bed exit, a beep sounds and the raise bed” indicator flashes, and I’m
unable to arm the system. Why?
If the bed is equipped with IntelliDrive® XL Transport System, you cannot arm bed exit when
the bed is in transport mode. To remove the bed from the transport mode, on the caregiver
control pod, press and hold the raise bed” control until you hear a single beep confirmation.
You can now arm the bed exit system.
5. I’m having difficultly moving the bed side to side after returning from a transport. Why?
If the bed is equipped with IntelliDrive® XL Transport System, while the transport wheel is on
the floor you will not be able to make small lateral movements. Raise the height of the bed to
remove from transport mode, in order to make small lateral movements or to move the bed
side to side. You can raise the height of the bed by disengaging transport mode by pressing
the “Disengage” control on the transport pod, or you can use raise the height of the bed using
the caregiver controls.
+
6. I’ve forgotten to retract the width of the bed before transport, can I adjust the width of
the bed when it’s unplugged?
Yes, but you will have to adjust it manually. Powered width expansion control only works
when plugged into power. To manually retract the bed, first use the CPR deflate to release
enough air from the surface. Then adjust the frame by pulling release lever on the head
siderail and simultaneously push slide extender inward. Next complete the same action on the
intermediate siderail. Repeat on other side of the bed.
Compella™ Bariatric Bed
Troubleshooting Guide
Question and Answer IconNeed help?

Do you have a question about the Hill-Rom Compella and is the answer not in the manual?

Summary

Troubleshooting Bed Width Adjustment

Continuous Beep After Width Adjustment

Bed width adjustment requires a single beep confirmation to indicate full adjustment.

Verify Power Connections for Width Adjustment

Ensure both bed and surface power cords are plugged in for powered width adjustment.

Check Communication Cable for Width Adjustment

Verify the communication cable between the surface blower box and frame is secure.

Troubleshooting Transport Mode Features

Weighing Patient in Transport Mode

Weighing is disabled in transport mode. Exit transport mode by holding "raise bed".

Arming Bed Exit in Transport Mode

Bed exit arming is disabled in transport mode. Exit transport mode to enable.

Lateral Movement After Transport

Small lateral movements are restricted in transport mode. Raise bed to enable.

Manual Bed Adjustments

Manual Width Adjustment When Unplugged

Manually adjust width using CPR deflate and side rail release levers.

Bed Operation and Patient Care

Adjusting Bed Length for Elevator Access

Use "Foot Shorter" control to shorten the bed length. Blower box removal is optional.

Lowering Bed for Patient Transfer

Use seat deflate feature or transport mode to lower the bed for easier patient access.

Mattress Inflation Troubleshooting

Ensure CPR mechanism is closed and use Max Inflate to re-inflate the mattress.

Programming New Patient Surface Settings

Program patient height and weight via the blower box control panel for automatic surface adjustment.

Battery Charging and Management

Plug bed into power when possible. Monitor battery indicators for bed and IntelliDrive® XL.

Summary

Troubleshooting Bed Width Adjustment

Continuous Beep After Width Adjustment

Bed width adjustment requires a single beep confirmation to indicate full adjustment.

Verify Power Connections for Width Adjustment

Ensure both bed and surface power cords are plugged in for powered width adjustment.

Check Communication Cable for Width Adjustment

Verify the communication cable between the surface blower box and frame is secure.

Troubleshooting Transport Mode Features

Weighing Patient in Transport Mode

Weighing is disabled in transport mode. Exit transport mode by holding "raise bed".

Arming Bed Exit in Transport Mode

Bed exit arming is disabled in transport mode. Exit transport mode to enable.

Lateral Movement After Transport

Small lateral movements are restricted in transport mode. Raise bed to enable.

Manual Bed Adjustments

Manual Width Adjustment When Unplugged

Manually adjust width using CPR deflate and side rail release levers.

Bed Operation and Patient Care

Adjusting Bed Length for Elevator Access

Use "Foot Shorter" control to shorten the bed length. Blower box removal is optional.

Lowering Bed for Patient Transfer

Use seat deflate feature or transport mode to lower the bed for easier patient access.

Mattress Inflation Troubleshooting

Ensure CPR mechanism is closed and use Max Inflate to re-inflate the mattress.

Programming New Patient Surface Settings

Program patient height and weight via the blower box control panel for automatic surface adjustment.

Battery Charging and Management

Plug bed into power when possible. Monitor battery indicators for bed and IntelliDrive® XL.

Overview

The Compella™ Bariatric Bed is a specialized medical bed designed to accommodate bariatric patients, offering a range of features to enhance patient comfort, safety, and caregiver efficiency. Its primary function is to provide a robust and adaptable support surface for larger individuals, while also incorporating advanced mobility and patient management capabilities.

Function Description:

The Compella™ Bariatric Bed serves as a comprehensive patient support system, capable of adjusting its dimensions and support surface properties to meet the unique needs of bariatric patients. It facilitates patient care by offering powered adjustments for bed width, length, and height, as well as specialized surface features like low air loss and pressure redistribution. The bed is designed to be highly maneuverable, especially when equipped with the IntelliDrive® XL Transport System, which aids in moving patients within healthcare facilities. Integrated weighing capabilities allow for accurate patient weight monitoring, crucial for medication dosing and care planning. Additionally, the bed incorporates safety features such as bed exit alarms to prevent falls and ensure patient security.

Usage Features:

  • Width Adjustment: The bed offers powered width adjustment, allowing it to expand or retract to either 40" or 50". This feature is crucial for accommodating patients of varying sizes and for navigating through doorways. A continuous beep signals that the bed is not fully adjusted, with a single beep confirmation indicating a complete adjustment. If the bed is unplugged, manual width adjustment is possible by using the CPR deflate feature to release air from the surface, then pulling release levers on the head and intermediate siderails while simultaneously pushing the slide extender inward.
  • Length Adjustment: The foot section of the bed can be shortened by pressing and holding the "Foot Shorter" control, which is useful for fitting the bed into elevators or smaller spaces. The surface blower box can also be removed to further reduce the bed's length by disconnecting hoses and lifting the box.
  • Height Adjustment and Patient Access: The bed's height can be adjusted to facilitate patient ingress and egress. For patients needing to get their feet closer to the ground, the low air loss surface offers a "seat deflate" feature, activated via the blower box by pressing "mattress adjust" and then "deflate seat." If equipped with IntelliDrive® XL, placing the bed in transport mode can lower it an additional 1.5 inches, further aiding patient access.
  • IntelliDrive® XL Transport System: This system enhances the bed's mobility, allowing for easier transport of patients. When in transport mode, certain functions like patient weighing and bed exit arming are disabled, indicated by a flashing "raise bed" indicator and a beep. To exit transport mode, caregivers must press and hold the "raise bed" control on the caregiver control pod until a single beep confirmation is heard. This system also affects lateral movement; if the transport wheel is on the floor, small lateral movements are restricted. Raising the bed height or disengaging transport mode via the transport pod's "Disengage" control enables lateral movement.
  • Patient Weighing: The bed includes an integrated weighing system. However, weighing is not possible when the bed is in transport mode. Exiting transport mode by pressing and holding the "raise bed" control will enable the weighing function.
  • Bed Exit System: The bed exit system can be armed to alert caregivers if a patient attempts to leave the bed. Similar to weighing, this system cannot be armed when the bed is in transport mode and requires exiting transport mode first.
  • Mattress Inflation and Surface Adjustment: If the mattress does not remain inflated, caregivers should check if the CPR mechanism is fully closed by turning it counter-clockwise. After ensuring it's closed, pressing "Max Inflate" on the air supply unit will quickly inflate the mattress. For new patients, it's essential to program their height and weight into the system via the blower box. This involves using arrow controls to navigate to height and weight settings, then using increase/decrease controls to input the patient's data. The surface will then automatically adjust pressure redistribution based on these inputs.
  • Battery Management: The bed and IntelliDrive® XL system are powered by batteries, which should be charged by plugging the bed into AC power whenever possible. The caregiver control panel and transport pod display battery charge indicators. For transport, it's recommended to have at least one bar on the transport pod's battery indicator. If the IntelliDrive® XL battery fully discharges during transport, the bed's height can be raised using caregiver controls to lift the transport wheel off the ground, allowing the bed to be connected to AC power.

Maintenance Features:

  • Communication Cable Verification: If the powered width adjustment is not working despite the bed and surface being plugged in, caregivers should verify that the communication cable between the surface blower box and the frame is securely connected.
  • CPR Mechanism Check: In cases of mattress deflation, ensuring the CPR mechanism is fully closed is a key troubleshooting step.
  • Power Cord Connection: Regular verification that both power cords for the bed and surface are securely plugged into power is essential for the proper functioning of powered features.
  • Battery Charging: Consistent charging of the bed's batteries by keeping it plugged into AC power when not in use is crucial for maintaining operational readiness, especially for the IntelliDrive® XL Transport System.
  • Manual Adjustments: The provision for manual width adjustment when the bed is unplugged serves as a practical maintenance and operational workaround, requiring specific steps involving air release and lever manipulation.

Hill-Rom Compella Specifications

General IconGeneral
CategoryBariatric Hospital Bed
Trendelenburg/Reverse TrendelenburgYes
Integrated ScaleYes
Power Requirements120V AC, 60 Hz
Battery BackupYes
Weight Capacity1000 lbs
Bed Width42 inches (107 cm)
Bed Length84 inches (213 cm)
Side RailsYes

Related product manuals