SERVICE MANUAL
Basic Care™
Bed
From Hill-Rom
Product No. P1440/P1441
For Parts or Technical Assistance man336
USA 1-812-934-1796 Canada (800) 267-2337
International: Contact your distributor.

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom Basic Care P1440 and is the answer not in the manual?

Overview

This is a service manual for the Basic Care™ Bed from Hill-Rom, specifically covering product numbers P1440 (electric model) and P1441 (manual model). It provides instructions for operation, maintenance, and includes parts lists for replacement components.

Function Description:

The Basic Care™ Bed is designed to provide patient support and facilitate caregiving by allowing adjustments to various sections of the sleep deck and overall bed height. The electric model utilizes linear actuators and electronic controls for motion, while the manual model employs cranks and mechanical mechanisms.

The bed's articulating sleep deck comprises head, seat, knee, and foot sections. The head, knee, and foot sections can articulate to change their angular orientation relative to the intermediate frame. Siderails are present on both sides of the bed, capable of being raised and lowered, and tucking below the sleep deck when lowered.

The electric model features low-voltage controls on the head siderails and foot end control panel for adjusting the head and knee sections' angles and the bed's height. It also includes a battery backup system that allows operation of CPR, Trendelenburg, and Reverse Trendelenburg functions for one cycle with a safe working load of 450 lb (204 kg) when no mains power is supplied. Lockout controls on the electric model can disable individual or all three motors.

Both electric and manual versions offer manually operated Trendelenburg and Reverse Trendelenburg functions. For the electric model, the hilow motor is used to achieve these positions after manual operation. A CPR release mechanism is also integrated, allowing for rapid flattening of the head section.

Important Technical Specifications:

Physical Description:

  • Head section slope (maximum): 72.5° ±1°
  • Knee section slope (maximum): 25° ±1°
  • Sleep deck height range: 18.3" to 30.1" ± 1" (46.5 cm to 76.5 cm ± 2.5 cm)
  • Trendelenburg position (maximum): 12° (+2°, -0°)
  • Reverse Trendelenburg position (maximum): 12° (+2°, -0°)
  • Bed lift capacity (maximum safe working load): 450 lb (204 kg)
  • Foot section lift capacity (maximum): 200 lb (91 kg)
  • Head section lift capacity (maximum): 200 lb (91 kg)
  • Maximum height of seat section (in Trendelenburg position): 23.5" (59.7 cm)
  • Siderail opening size: 3.875" (9.843 cm)
  • Distance between siderails: 11.375" (28.893 cm)
  • Length—with standard sleep deck: 89" to 91" (226 cm to 231 cm)
  • Length—with short sleep deck: 86" to 88" (218 cm to 224 cm)
  • Sleep deck length—standard model: 79" to 81" (201 cm to 206 cm)
  • Sleep deck length—short model: 76" to 78" (193 cm to 198 cm)
  • Maximum width—with siderails raised: 43.7" ± 0.8" (110.9 cm ± 2.0 cm)
  • Minimum width—with siderail stored: 36.6" ± 0.8" (93.0 cm ± 2.0 cm)
  • Sleep deck width: 36.25" ± 0.5 (92.08 cm ± 1.3 cm)
  • Maximum headboard height: 44" (112 cm)
  • Minimum underbed clearance: 5.5" (13.9 cm)
  • Wheel base: 61" (155 cm)
  • Caster size: 5" (13 cm)
  • Total weight: 310 lb (141 kg)
  • Mattress width: 36" to 37" (91 cm to 94 cm)
  • Mattress length—standard model: 80" to 81" (203 cm to 206 cm)
  • Mattress length—short model: 76" to 77" (193 cm to 196 cm)
  • Mattress thickness: 6" to 7" (15 cm to 18 cm)

Environmental Conditions for Use:

  • Temperature: 50°F to 104°F (10°C to 40°C) ambient temperature
  • Relative humidity (RH): 30% to 95%, non-condensing
  • Atmospheric pressure: 700 hPa to 1060 hPa

Electrical Description:

  • Rated voltage: 100 V AC, 110 V AC to 115 V AC, 120 V AC to 127 V AC, 220 V AC to 230 V AC, 240 V AC
  • Power/Input: 100 V AC to 127 V AC—1.0 A, 220 V AC to 240 V AC—0.5 A
  • Frequency: 50/60 Hz

Battery Specifications:

  • Maximum battery life (no functions operated, bed disconnected from power source): 24 hours
  • Time necessary to recharge a fully discharged battery: 8 hours
  • Battery fuse: 15A
  • Maximum hilow cycles with fully charged battery:
    • With 0 lb (0 kg) on bed: 21 cycles
    • With 250 lb (113 kg) on bed: 15 cycles
    • With 450 lb (204 kg) on bed: 11 cycles

Regulations, Standards, and Codes:

  • Technical and Quality Assurance Standards: UL 60601-1, CSA® C22.2 No. 601.1, EN 60601-2-38 (including amendment 1), EN 60601-1, IEC 60601-1-2, EN ISO 9002
  • Equipment classification per EN 60601-1: Class I equipment, internally powered equipment
  • Degree of protection against electric shock: Type B
  • Classification according to Directive 93/42/EEC: Class I
  • Degree of protection against the presence of flammable anaesthetic mixtures: Not for use with flammable anaesthetics.
  • IPX classification: IPX 4—Protected against unpressurized spraying and splashing water.

Usage Features:

The bed is available in both electric (P1440) and manual (P1441) models. The electric model offers convenient electronic controls for patient and caregiver adjustments of the head and knee sections, as well as bed height. The manual model relies on cranks for these adjustments. Both models incorporate Trendelenburg and Reverse Trendelenburg capabilities. The electric model's battery backup ensures critical functions remain operational during power outages.

Maintenance Features:

The manual outlines detailed troubleshooting procedures, starting with initial actions and progressing through function checks to isolate problems. It provides step-by-step instructions for removal, replacement, and adjustment of various components, including:

  • Control Box Assembly
  • Foot End Control Panel
  • Trendelenburg Control Box
  • Trendelenburg or Reverse Trendelenburg Assembly
  • Trendelenburg and Reverse Trendelenburg Indicators
  • Fuses
  • Battery Backup
  • Head Motor
  • CPR Cable
  • Knee Motor
  • Hilow Motor
  • Manual Crank Handle Assembly
  • Head or Knee Crank Rod and Drive Screw Assembly (Manual Model Only)
  • Hilow Crank Rod and Drive Screw Assembly Including Gas Springs (Manual Model Only)
  • CPR Handle Assembly
  • Siderails
  • Siderail Slide Bracket
  • Siderail Retraction Assembly
  • Siderail Control Assembly
  • Siderail Control Cable
  • Siderail Top Cane
  • Siderail Wire Cover
  • Central Brake and Steer—Caster Assemblies
  • Central Brake and Steer—Pedal Assemblies
  • Casters for Individual Brake and Steer
  • Bumper
  • Sleep Surface Panels

The manual also includes general procedures for cleaning and care, emphasizing the use of appropriate disinfectants and avoiding harsh cleansers. Specific instructions for handling electronic components like P.C. boards are provided, highlighting the importance of antistatic precautions. Lubrication requirements are detailed, specifying suitable lubricants for various parts, including Oilite® bearings and motor gears. A comprehensive preventive maintenance schedule and checklist are included to guide annual maintenance and testing, ensuring a long operative life and compliance with JCAHO requirements. The bed also has a "Sleep Mode" procedure to conserve battery power when disconnected from AC for extended periods.

Hill-Rom Basic Care P1440 Specifications

General IconGeneral
BrandHill-Rom
ModelBasic Care P1440
CategoryMedical Equipment
LanguageEnglish