2.14 The CPR Function Does Not Operate
Chapter 2: Troubleshooting Procedures
Basic Care™ Bed Service Manual (man336) Page 2 - 25
2
2.14 The CPR Function Does Not Operate
1. The CPR cable (A) is connected to the CPR handle (B) (see figure 2-2 on
page 2-26)
Yes No
↓→Connect the CPR cable (A) to the CPR handle (B). If this
corrects the problem, go to “Final Actions” on page 2-4.
Otherwise, go to step 2.
2. The CPR wire (C) is held tightly in the bracket (D).
Yes No
↓→Tighten the screw (E) that attaches the CPR wire (C) under the
bracket (D). If this corrects the problem, go to “Final Actions”
on page 2-4. Otherwise, go to step 3.
3. The CPR cable (A) and CPR wire (C) are connected to the head motor (F).
Yes No
↓→Attach the CPR cable (A) and/or the CPR wire (C) to the head
motor (F). If this corrects the problem, go to “Final Actions” on
page 2-4. Otherwise, go to step 4.
4. Replace the head motor (F). If this corrects the problem, go to “Final
Actions” on page 2-4. Otherwise, call Hill-Rom Technical Support at
1-812-934-1796.