Lubrication Requirements
Chapter 6: General Procedures
Basic Care™ Bed Service Manual (man336) Page 6 - 3
6
When servicing the P.C. board do the following:
Make sure hands are clean and free of moisture, oily liquids, etc.
Only handle the P.C. board by its outer edges.
Do not touch the P.C. board components. Finger contact with the board
surface and/or with its components can leave a deposit that will result in
board (and component) deterioration.
When working with electronics, wear an appropriate antistatic strap, and
make sure it is correctly grounded.
Service the removed P.C. board at a static-free workstation that is correctly
grounded.
For shipping and storage, place the removed P.C. board in an antistatic
protective bag.
Lubrication Requirements
WARNING:
Obey the product manufacturer’s instructions. Failure to do so could
result in personal injury or equipment damage.
CAUTION:
Do not use silicone-based lubricants. Equipment damage could occur.
Oilite®
1
bearings and bushings are utilized in several places on the system. By
retaining oil, the pores give a self-lubricating quality to the bearings and
bushings. If any silicone-based lubricant is applied to the bearings and
bushings or anywhere else on the system, this self-lubricating quality is
neutralized.
It is safe to apply the following lubricants to the system (see table 6-1 on page
2-3):
Table 6-1. Lubricants
1. Oilite® is a registered trademark of Beemer Precision, Incorporated.
Part Number Description
8252 (1400) Non-detergent SAE30 weight oil (apply to Oilite®
bearings and bushings)
SA3351 (1400) 4 oz lithium grease