Specifications
Chapter 1: Introduction
Basic Care™ Bed Service Manual (man336) Page 1 - 5
1
Table 1-6. Battery Specifications
Regulations, Standards, and Codes
Table 1-7. Regulations, Standards, and Codes
Condition Range
Maximum battery life, with no functions
operated and with the bed disconnected
from its power source
24 hours
Time necessary to recharge a fully dis-
charged battery
8 hours
Battery fuse 15A
Maximum hilow cycles with fully charged battery:
With 0 lb (0 kg) on bed 21
With 250 lb (113 kg) on bed 15
With 450 lb (204 kg) on bed 11
Classification Standard
Technical and Quality Assur-
ance Standards
UL 60601-1
CSA®
a
C22.2 No. 601.1
EN 60601-2-38, including amendment 1
EN 60601-1
IEC 60601-1-2
EN ISO 9002
a. CSA® is a registered trademark of Canadian Standards Association, Inc.
Equipment classification per EN
60601-1
Class I equipment, internally powered equipment
Degree of protection against
electric shock
Type B
Classification according to
Directive 93/42/EEC
Class I
Degree of protection against the
presence of flammable anaes-
thetic mixtures
Not for use with flammable anaesthetics.
IPX classification IPX 4—According to IEC 60529, rating for protection
against fluid ingress and identified as equipment that is
protected against unpressurized spraying and splashing
water.