EasyManua.ls Logo

Hill-Rom Basic Care P1440 - Fuses; Removal

Hill-Rom Basic Care P1440
160 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.6 Fuses
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 14 Basic Care™ Bed Service Manual (man336)
4.6 Fuses
Tools required: Phillips head screwdriver
Needle nose pliers
Removal
1. Raise the head section to the highest position.
2. Adjust the bed height to have access to the cover (A) (see figure 4-6 on
page 4-14).
Figure 4-6. Control Box Assembly
SHOCK HAZARD:
Disconnect the bed from its power source. Failure to do so could result
in personal injury or equipment damage.
3. Disconnect the bed from its power source.
4. Remove the six screws (C) attaching the cover (A) to the control box
assembly (B).

Table of Contents