Page 50 171051(9) - Hill-Rom® 900 Electric Bed User Manual
PROPRIETARY AND CONFIDENTIAL DRAFT: 18 May 2018
To activate the Bed exit alert detection:
Activating the bed exit detection to a given degree of sensitivity is subject to the
following pre-conditions that guarantee effective patient detection.
Pre-conditions for the activation of the Position and Exiting modes:
• The patient is in the center of the bed and aligned with the hip position
markers.
If the pre-conditions for activation are not met, a beep sounds. In this case,
follow the instructions and repeat the procedure.
Only one bed exit mode can be activated at a time.
1. Press the button of the required function and hold it until
activation (the green light flashes during activation).
2. A beep sounds and the light turns green to indicate that the
function has been activated.
If the activation fails, a beep sounds three times and the light does
not come on.
To deactivate the Bed exit alert monitor
Press the button of the required function and the green light goes off.
When an alert sounds
When the bed exit monitor is active and it detects alert conditions, a
continuous alert signal sounds, the green light corresponding to the
function flashes, the night light comes on and a signal is sent to the
duty nurse (see “Sending bed exit alerts” page 51).
Press the “Alert mode suspension” button to deactivate the alert for 30
seconds or 5 minutes (see “Selecting the duration of the alert
suspension” page 51).
If a load weighing more than 9 kg is added or removed, it is necessary to
proceed with an initialization.
Position
Exiting Out of bed