Decontamination
Maintenance
Maintenance
Safety recommendations
Only facility-authorized personnel should perform maintenance of the Hill-
Rom® 900 bed.
Before maintenance or servicing works:
• ensure that the bed has been immobilized (if no movements are required),
• lock out all electrical functions,
• disconnect the bed from the mains if no electrical operations are planned,
• secure the sleep surface and take whatever steps are necessary to prevent
any movement.
• Do not work on the bed when it is occupied.
All devices connected to the CAN socket, used exclusively for maintenance
operations, must meet the requirements of IEC 60950-1.
Never open or pierce an electric actuator.
For all problems with actuators (e.g., blockage), contact our after-sales service.
Preventive maintenance
A service manual and a catalog of spare parts are supplied on delivery, but can
also be obtained on demand from Hill-Rom After-Sales. Hill-Rom guarantees
that the original functional parts or parts performing equivalent functions will
remain available for 7 years after the corresponding range goes out of
production.
The product test plan is based on 10 years of normal use.
The frequency of inspections must be adapted to the general condition of the
product and it use, for example, if the bed is used by heavy patients. It is the
responsibility of the facility to implement a preventive maintenance program for
the bed's functions under its conditions of use.