187692(10) - EN - P280 Mattress Instructions for Use Page 9
Destination
Specifications
Control Unit
Front and Back Views of the P280 Control Unit
The control unit may generate noise and vibration that can cause discomfort
to the patient.
Characteristics Values
Model
M04566
M04584
Dimensions 29 x 20 x 12 cm / 11.4" x 8" x 4.6"
Weight 2.2 kg / 5 lb
Power supply 220-240 V~, 50/60 Hz 120 V~, 50/60 Hz
Rated power
0.07 A @ 220-240 Vac 0.17 A @ 120 Vac
Maximum energy consumed 5.2 Wh
Operation of the device Continuous
Sleeve material ABS PC
Device: sound pressure level /
sound power (ISO 3744)
45 dB(A) / 61 dB(A)
Fuses T1AL, 250V
Compression 0-10,5 kPa
Compressor flow rate 10 l/min
IEC 60601-1 Classification Classe II
Degree of protection provided by
the cover (IEC 60529)
IP41: Protected against access to hazardous
parts with a wire and against vertically falling
water drops
Protection against inflammable
anesthetic mixtures
Not for use with flammable anesthetics.
Control Panel
Hose Connector
Fuses
Filter
Main Switch
Power Socket
Handle Attachment