187692(10) - EN - P280 Mattress Instructions for Use Page 21
Installing
the Patient
6. Adjust the length of the straps to safely
secure the mattress.
7. Make sure that the mattress is correctly
installed,securely attached, centered on the
mattress support platform and pushed tightly
against the footboard to avoid the danger of
entrapment.
8. Make sure that the attachments do not
become entangled in the moving parts of the
mattress support platform, such as actuators,
CPR handles, etc. With articulated frames,
make sure that the straps of the mattress are only attached to the mobile
head and foot sections and NOT to the main frame.
9. Attach the control unit to the footboard of the
bed using the handle aAttachment.
10. Connect the hose to the control unit.
The connector clicks when correctly in
position
.
11. Make sure that the hose cable does not
become entangled in the barriers of the bed.
Where appropriate, place the cable on the
handrail panel footboard.
12. Connect the power cord to the control unit.