8
―――――――――――――――――――――――――――
Chapter 2 Specifications
――――――――――――――――――――――――
Battery Life More than 500 times (R6P in use) or
more than 1400 (LR6 in use)
(30-second measurement/30-second
pause cycle)
Overvoltage
protection
250 VAC for one minute
between E-P (S) and E-C (H) terminals
Insulation resistance 100 MΩ or more/500 V DC
between electric circuit and case
Dielectric voltage 3000 V AC for one minute
between electric circuit and case
External dimensions Approx. 164W x 119H x 88D mm
(6.46W" x 4.37H" x 3.47D")
(excluding protrusion)
Mass Approx. 800 g (28.2oz.) (instrument only)
Accessories 9214 AUXILIARY EARTHING ROD
9215 MEASURING CABLE
(black 5m, yellow 10 m,
red 20 m each one,
9216 CABLE WINDER x 3)
9393 CARRYING CASE
R6P manganese battery x 6
Instruction manual, Hand strap
Option 9050 EARTH NETS 30 x 30cm
Standards applying
Earthing resistance
measurement
Safety
EMC
Environment
protection
JIS C1304-1995
EN61557-5:1997
JIS C1004-1996
EN61010-1:2001
Pollution Degree 2, Measurement
Category II (anticipated transient
overvoltage 500 V)
EN61010-2-031:1994
EN61326:1997+A1:1998+A2:2001
EN60529:1991 "IP40"
JIS C920-1993