EasyManua.ls Logo

Hioki 3151 - Measurement Range and Tolerances

Hioki 3151
54 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
―――――――――――――――――――――――――――
Chapter 2 Specifications
――――――――――――――――――――――――
Accuracy is guaranteed for 1 year at following conditions.
(Temperature and humidity: 23 5 (73 5 ), 80%
RH or less)
Measurement
item
Measurement range Tolerances
Earthing
resistance
()
10 (0 to 11.5)
100 (0 to 115)
1000 (0 to 1150)
2.5%f.s.
Earth voltage (V) 30 (0 to 30) 3.0%f.s.
When using 2-pole method, applies 100 and 1000 only
Effect of temperature Within 0.1%/
at 0 to 40 (32 to 104 )
Effect of auxiliary
earth resistance
Within 5% when fluctuation is
0to5k
Effect of earth
voltage
Within 2% at 0 to 5 V
Within 2% at 0 to 10 V
(at 50 or 60 Hz)
Within 5% at 0 to 3 V
(at DC, 16 2/3, 400 Hz)
Effect of power
voltage
Within specifications for 6 - 10 V DC
Effect of radiated
radio-frequency
electromagnetic field
At 3 V/m
Within +2.5%f.s. to -40%f.s.
(earthing resistance)
2.2 Measurement Range and
Tolerances