25
Chapter 5 Instructions of Command Sets
SYSTem:BLIT <Boolean>
SYSTem:BLIT:TIME <second>
SYSTem:AOFF:TIME <minute>
SYSTem:CPERcent < 0-20 | 4-20 >
SYSTem:BATTery?
SYSTem:DEFAult
Description Enable or disable the back lit.
Parameter Value to be entered to <Boolean>: ON, 1: OFF, 0
The backlit will be auto turned off depending on below time setting. To turn
off backlit by this command once the time is set to "0".
Description Set the time of the back lit.
Parameter Value to be entered to <second>: 0 to 99
(Set "0" to disable auto turning off backlit.)
Stored in non-volatile memory.
Description Set the time of the auto power save function.
Parameter Value to be entered to <minute>:0 to 99
(Set "0" to disable auto power save function.)
Stored in non-volatile memory.
Description
Selects the % conversion display (4-20mA or 0-20mA) for DCmA
measurements. It is available when % conversion display is enabled
by the CONF:CURR:PERC command.
Parameter
0-20: 0 mA (0 %) 20 mA (100 %)
4-20: 4 mA (0 %) 20 mA (100 %)
Example
SYST:CPER 4 to 20
Stored in non-volatile memory.
Description Query the
battery level.
Parameter
<NR3>Outputs a formatted text string. Response from 0 to 100%
corresponds to battery voltage of 6.0 to 10.0 V.
Example
0% (6.0 V) to 100% (10 V)
Stored in non-volatile memory.
Description
Returns meter settings to their factory defaults. Internal adjustments
are unaffected. The meter resets after executing this command.