Chapter 5 Instructions of Command Sets
14
This subsection describes the SCPI subsystem commands for the
meter. The meter only accepts the UPPER CASE part of command. It
is unnecessary to send complete command characters.
CONFigure Subsystem:
CONFigure
:CURRent
:DC [<range>]
:AC [<range>]
:ACDC [<range>]
:PERCent
5.2 SCPI Commands
Some SCPI Symbol Conventions
TextSymbol
Meaning
[ ]
Option: can be omitted
|
Exclusive OR
< >
Defined element
( )
Comment
?
Question mark
:
Separated two command
keywords
;
Compound commands
Description Set to DC, AC or AC+DC current measurement.
(Only 3801-50: AC+DC)
The function switch is at current measuring.
When measuring DCmA, select the % conversion. Use the
SYST:CPER command to select the 4-20 mA or 0-20 mA
scale.
Parameter If parameter is omitted, the meter is set to auto range.
Enter the current measurement range to <range>.
The function switch is at μA measuring: 500
u, 5000u
500
u(510.00 μA), 5000u(5100.0 μA)
The function switch is at mA.A measuring: 0.05, 0.5, 5 or 10
0.05(51.000 mA), 0.5(510.00 mA), 5(5.1000 A), 10(10.000 A)
Example
CONF:CURR:DC : Set to DC current.
CONF:CURR:AC 0.5 : Set to AC current and 510.00 mA
range.
CONF:CURR:PERC : For DCmA measurements, select the
% conversion..