15
Chapter 5 Instructions of Command Sets
CONFigure
:VOLTage
:DC [<range>]
:AC [<range>]
:ACDC [<range>]]
CONFigure
:FCOUnter
:PREScale<1 / 100>
CONFigure:FREQuency
[<range>]
Description Set to DC, AC or AC+DC voltage measurement. (Only 3801-
50: AC+DC)
The function switch is at mV measuring or V measuring.
Parameter • If parameter is omitted, the meter is set to auto range.
• Enter the voltage measurement range to <range>:
The function switch is at mV measuring : 0.05 0.5 or 1
0.05(51.000 mV), 0.5(510.00 mV), 1(1000.0 mV)
The function switch is at V measuring: 5, 50, 500 or 1000
5(5.1000 V), 50(51.000 V), 500(510.00 V), 1000(1000.0 V)
Example
CONF:VOLT:DC : Set to DC voltage measuring.
CONF:VOLT:AC 0.5 : Set to AC voltage and 510 mV
range.
Description Select frequency counter prescaling. (Only 3801-50)
The function switch is at
diode measuring. Range is auto range.
Parameter 1 :Disables frequency counter prescaling (divisor = 1).
100 :Disables frequency counter prescaling (divisor = 100).
Example
CONF:FCOU:PRES 1: Disables frequency counter
prescaling (divisor = 1).
Description Frequency measurement can be selected for voltage or current
measurement.
The function switch is at voltage or current measuring.
The display shows both main and sub displays simultaneously.
The frequency range changes when measuring frequency.
(Only 3801-50)
The function switch is at D(Hz).
Parameter • If parameter is omitted, the meter is set to auto range.
• Enter the resistance measurement range to <range>. (Unit:
Hz)
: 100, 1000, 10K, 100K, 1000K, 10M
∗, or 100M∗(Only
frequency measurement)
∗ Only 3801-50 frequency counter measurement.
Example
CONF:FREQ 10k : Set to frequency measuring and 9.9999
kHz range.