Chapter 5 Instructions of Command Sets
16
CONFigure
:PULSe
:PWIDth [<range>]
:NWIDth [<range>]
:PDUTycycle
:NDUTycycle
CONFigure:RESistance [<range>]
CONFigure:CONTinuity [<range>]
CONFigure:CONDuctance
CONFigure:CAPacitance [<range>]
Description Set to pulse width measurement, pulse duty measurement.
This can be selected when voltage, current or frequency
counter∗ (divisor = 1) measurement is selected.
∗ Only 3801-50
Parameter • If parameter is omitted, the meter is set to auto range.
• Enter the pulse measurement range to <range>:
• Set the parameter to 5 to select the 1999.9 ms range for
pulse width measurement.
:0.5, 5
0.5(510.00 ms), 5(1999.9 ms)
Example
CONF:PULS:NWID 0.5 :Sets negative pulse width
measurement in the 510.00 ms range.
CONF:PULS:NDUT :Measures negative pulse duty.
Description Set to 2-wire resistance measuring.
The function switch is at resistance measuring.
Parameter • If parameter is omitted, the meter is set to auto range.
• Enter the resistance measurement range to <range>. (Unit:
Ω): 500, 5K, 50K, 500K, 5M, 50M or 500M
∗
∗
Only 3801-50
Example
CONF:RES 50k : Set to resistance measuring and 51.000 kΩ
range.
Description Set to 2-wire resistance measuring with the continuity test.
The function switch is at resistance measuring.
Parameter • If parameter is omitted, the meter is set to auto range.
• Enter the resistance measurement range to <range>.(Unit:
Ω): 500, 5K, 50K, 500K, 5M, 50M or 500M
∗.
∗Only 3801-50
Example
CONF:CONT 500 : Set to resistance measuring with the
continuity test and 510.00 Ω range.
Description • Set to conductance measuring . The function switch is at
resistance measuring.
Example •
CONF:COND : Set to conductance measuring.
Description Set to capacitance measurement . The function switch is at C
measuring.
Parameter If parameter is omitted, the meter is set to auto range.
Capacitance measurement range to <range>.(Unit: F)
: 10n, 100n, 1000n, 10u, 100u, 1000u, 10m or 100m
Example
CONF:CAP 100u: Set to capacitance measurement and
99.99
μF range.