Chapter 5 Instructions of Command Sets
16
CONFigure
:PULSe
:PWIDth [<range>]
:NWIDth [<range>]
:PDUTycycle
:NDUTycycle
CONFigure:RESistance [<range>]
CONFigure:CONTinuity [<range>]
CONFigure:CONDuctance
CONFigure:CAPacitance [<range>]
Description Set to pulse width measurement, pulse duty measurement.
This can be selected when voltage, current or frequency
counter (divisor = 1) measurement is selected.
Only 3801-50
Parameter If parameter is omitted, the meter is set to auto range.
Enter the pulse measurement range to <range>:
Set the parameter to 5 to select the 1999.9 ms range for
pulse width measurement.
:0.5, 5
0.5(510.00 ms), 5(1999.9 ms)
Example
CONF:PULS:NWID 0.5 :Sets negative pulse width
measurement in the 510.00 ms range.
CONF:PULS:NDUT :Measures negative pulse duty.
Description Set to 2-wire resistance measuring.
The function switch is at resistance measuring.
Parameter If parameter is omitted, the meter is set to auto range.
Enter the resistance measurement range to <range>. (Unit:
): 500, 5K, 50K, 500K, 5M, 50M or 500M
Only 3801-50
Example
CONF:RES 50k : Set to resistance measuring and 51.000 k
range.
Description Set to 2-wire resistance measuring with the continuity test.
The function switch is at resistance measuring.
Parameter If parameter is omitted, the meter is set to auto range.
Enter the resistance measurement range to <range>.(Unit:
): 500, 5K, 50K, 500K, 5M, 50M or 500M
.
Only 3801-50
Example
CONF:CONT 500 : Set to resistance measuring with the
continuity test and 510.00 range.
Description Set to conductance measuring . The function switch is at
resistance measuring.
Example
CONF:COND : Set to conductance measuring.
Description Set to capacitance measurement . The function switch is at C
measuring.
Parameter If parameter is omitted, the meter is set to auto range.
Capacitance measurement range to <range>.(Unit: F)
: 10n, 100n, 1000n, 10u, 100u, 1000u, 10m or 100m
Example
CONF:CAP 100u: Set to capacitance measurement and
99.99
μF range.