17
Chapter 5 Instructions of Command Sets
CONFigure:DIODe
CONFigure:TEMPerature <K | J>
CONFigure? [@2]
Description Set to diode measuring with continuity test.
The function switch is at diode measuring.
Description Set to temperature measuring with continuity test.
Only 3801-50 : J Termocouple
The function switch is at C measuring.
Paramete
K, J: Termocouple sensorType K or J.
Example CONF:TEMP K: Set to temperature of the K termocouple.
Description Query the function of the primary or secondary display.
Parameter If parameter is omitted, query the function of the main display.
Response Return <"function range,resolution"> format string, the
examples are as following:
Example Function Range Resolution
VOLT +5.000000E-01,+1.000000E-05
DCV
510.00
mV
0.01 mV
CURR:AC +5.000000E-02,+1.000000E-06
ACA
51.000
mA
0.001 mA
VOLT:DBM
dBm
CPER:0-20mA
+5.000000E-02,+1.000000E-06
DCmA
percentage disply:
0-20mA
51.000
mA
0.001 mA
FREQ +1.000000E+03,+1.000000E-02
Frequency
999.99
Hz
0.01 Hz
PULS:PWID +5.000000E+00,+1.000000E-04
Pulse
width(Positive)
1999.9
ms
0.1 ms
PULS:NDUT
Pulse duty
ratio(Negative)
CAP +1.000000E-08,+1.000000E-12 CCapacitance
9.999
nF
0.001 nF
RES+5.000000E+07,+1.000000E+03
R
Resistance
51.000
M
0.01 M
COND+5.000000E-07,+1.000000E-11
R
Conductance
510.00
nS
0.01 nS
CONT +5.000000E+02,+1.000000E-02
R
Continuous)
510.00
0.01
DIOD
Diode
TEMP:TC K CEL
Temperature
K-thermocouple