9
Operating Precautions
索引
4
3
2
1
7
6
5
10
9
8
The maximum rated voltage between input terminals
and ground is 600 V AC (150 V AC for model IR4030).
Attempting to measure voltages exceeding 600V AC
(150 VAC for model IR4030) with respect to ground
could damage the instrument and result in personal
injury.
1000 V or 600 V may be labeled depending on the sup-
plied test leads, but this is the rating of the test lead
and not the rating performance of the IR 4000series.
Please refer to the Specifications for the rating perfor-
mance of this instrument.
Before attaching to or removing the test lead from the
instrument, please remove the Test Lead from the
tested objected and turn the function selector to OFF.
Test leads should only be connected to the secondary
side of a breaker, so the breaker can prevent an acci-
dent if a short circuit occurs. Connections should never
be made to the primary side of a breaker, because
unrestricted current flow could cause a serious acci-
dent if a short circuit occurs.
Please only use batteries for electrical supply. Any
other electrical supply may damage the instrument and
tested object and cause electric shock.
Touching any of the high-voltage points inside the
instrument is very dangerous. Do not attempt to modify,
disassemble or repair the instrument; as fire, electric
shock and injury could result.
To avoid electric shock when measuring live lines, wear
appropriate protective gear, such as insulated rubber
gloves, boots and a safety helmet.