Chapter 2 Measurement Procedures
24
Ex. When measuring the voltage between circuit and ground
1.
Use the function selector to select the V function.
2.
Connect the black test lead to the ground side of the
object being measured.
3.
Connect the red test lead to the line to be measured.
4.
Read the value after the needle has stabilized.
1
2
3
4
Always connect the test lead
to the
secondary side of the
breaker.
EARTH
LINE
Caution:
MEASURE key.
Do not press
the
Load
(secondary side)
Source
(primary side)
During measuring, do not selector over to the
other function.
For waveforms other than sine waves, some
errors may occur.