43
Accuracy Table
索
引
6
5
4
3
2
1
PEAK MAX/PEAK MIN
Accuracy guarantee range (Resolution)
Accuracy guarantee
frequency range
Measurement accuracy
±10 A to ±1000 A (1 A) 10 Hz ≤ f < 45 Hz ±1.8% rdg.±7 A
45 Hz ≤ f ≤ 66 Hz ±1.3% rdg.±7 A
66 Hz < f ≤ 1 kHz ±2.0% rdg.±7 A
±1001 A to ±1500 A (1 A) 10 Hz ≤ f < 45 Hz ±2.3% rdg.±7 A
45 Hz ≤ f ≤ 66 Hz ±1.8% rdg.±7 A
66 Hz < f ≤ 1 kHz ±2.5% rdg.±7 A
(3) DC current
Measurement value/MAX/MIN/AVE
Accuracy guarantee range (Resolution) Measurement accuracy
±1.0 A to ±30.0 A (0.1 A) ±1.3% rdg.±0.8 A
±30.1 A to ±999.9 A (0.1 A) ±1.3% rdg.±0.3 A
PEAK MAX/PEAK MIN
Accuracy guarantee range (Resolution) Measurement accuracy
±10 A to ±1000 A (1 A) ±1.3% rdg.±7 A
±1001 A to ±1500 A (1 A) ±1.8% rdg.±7 A