Accuracy Table
(4) AC+DC current
Measurement value/MAX/MIN/AVE
Accuracy guarantee
range (Resolution)
Accuracy guarantee
frequency range
Measurement accuracy
Filter OFF Filter ON
1.0 A to 30.0 A (0.1 A) 10 Hz ≤ f < 45 Hz ±1.8% rdg.±1.2 A ±2.3% rdg.±1.2 A
DC, 45 Hz ≤ f ≤ 66 Hz ±1.3% rdg.±1.8 A ±1.8% rdg.±1.8 A
66 Hz < f ≤ 1 kHz ±2.0% rdg.±1.2 A –
30.1 A to 900.0 A (0.1 A) 10 Hz ≤ f < 45 Hz ±1.8% rdg.±0.7 A ±2.3% rdg.±0.7 A
DC, 45 Hz ≤ f ≤ 66 Hz ±1.3% rdg.±1.3 A ±1.8% rdg.±1.3 A
66 Hz < f ≤ 1 kHz ±2.0% rdg.±0.7 A –
900.1 A to 999.9 A (0.1 A) 10 Hz ≤ f < 45 Hz ±2.3% rdg.±0.7 A ±2.8% rdg.±0.7 A
DC, 45 Hz ≤ f ≤ 66 Hz ±1.8% rdg.±1.3 A ±2.3% rdg.±1.3 A
66 Hz < f ≤ 1 kHz ±2.5% rdg.±0.7 A –
PEAK MAX/PEAK MIN
Accuracy guarantee range (Resolution)
Accuracy guarantee
frequency range
Measurement accuracy
±10 A to ±1000 A (1 A) 10 Hz ≤ f < 45 Hz ±1.8% rdg.±7 A
DC, 45 Hz ≤ f ≤ 66 Hz ±1.3% rdg.±7 A
66 Hz < f ≤ 1 kHz ±2.0% rdg.±7 A
±1001 A to ±1500 A (1 A) 10 Hz ≤ f < 45 Hz ±2.3% rdg.±7 A
DC, 45 Hz ≤ f ≤ 66 Hz ±1.8% rdg.±7 A
66 Hz < f ≤ 1 kHz ±2.5% rdg.±7 A