45
Accuracy Table
6
5
4
3
2
1
(5) Current frequency/Voltage frequency
Range (Auto-range threshold)
Accuracy guarantee range
(Resolution)
Measurement accuracy
9.999 Hz (more than 9999 counts) 1.000 Hz to 9.999 Hz (0.001 Hz) ±0.1% rdg.±0.003 Hz
99.99 Hz (more than 9999 counts/
less than 900 counts)
1.00 Hz to 99.99 Hz (0.01 Hz) ±0.1% rdg.±0.01 Hz
999.9 Hz (less than 900 counts) 1.0 Hz to 999.9 Hz (0.1 Hz) ±0.1% rdg.±0.1 Hz
(6) INRUSH (Rush current)
INRUSH measurement value
Accuracy guarantee range (Resolution)
Accuracy guarantee
frequency range
Measurement accuracy
10.0 A to 999.9 A (0.1 A) DC, 20 Hz ≤ f ≤ 500 Hz ±5.0% rdg.±1.3 A
INRUSH PEAK value
Accuracy guarantee range (Resolution)
Accuracy guarantee
frequency range
Measurement accuracy
±10 A to ±1000 A (1 A) DC, 20 Hz ≤ f ≤ 500 Hz ±6.0% rdg.±10 A
±1001 A to ±1500 A (1 A) DC, 20 Hz ≤ f ≤ 500 Hz ±8.0% rdg.±10 A

Related product manuals